      KLD Mono&Multiple Water Quality Analyzer
      On-line Analyzer for Ammonia and Phosphate and Total Phosphorus Monitoring

      On-line Analyzer for Ammonia and Phosphate and Total Phosphorus Monitoring

          KLD-C-NH3-N & PO& TP is an automatic on-line analyzer independently developed by KLD. Using imported CPU, highly precise signal processor and simple but steady loop flow system to effectively reduce signal fluctuation and to ensure the accuracy and stability of data. Complete separation between electronics and hydraulics allows long-term & stable running. It is suitable for several types of water matrix like river water, surface water and industrial effluent.

      Hits:266
      Product Presentation

          KLD-C-NH3-N & PO4 & TP is an automatic on-line analyzer independently developed by KLD. Using imported CPU, highly precise signal processor and simple but steady loop flow system to effectively reduce signal fluctuation and to ensure the accuracy and stability of data. Complete separation between electronics and hydraulics allows long-term & stable running. It is suitable for several types of water matrix like river water, surface water and industrial effluent.


      Features

      1.      Robust & Reliable

      Designed for on-line industrial and environmental applications. Complete separation between electronics and hydraulics allows long-term stable operations.

      2.      Precise Reagent Dosage

      CPU controlled photoelectric dosing system for accurate reagent dosage. Simple & steady structure ensures long-term service life.

      3.      Easy Installation

      Delivered after a long & successful series of factory tests, it is ready for installation and setup. To start monitoring just need to connect reagent tubes, sample tube, water tube, waste tube and power supply, then set up the parameters.

      4.      Automatic Calibration

      The analyzer performs calibration according to the calibration interval and type set up by users. If the result exceeds selected limits, the original calibration parameters will not be updated and an alarm signal will be outputted.

      5.      Automatic Monitoring

      Measuring interval can be freely set up by users. Between two measurements the analyzer remains in standby.


      Benefits

      ?  Automatic monitoring system gives alarm signal in abnormal condition. Easy and low maintenance; Low operating cost.

      ?  4-20mA analog signal output; standard RS232, RS485 digital signal output, MODBUS protocol.

      ?  New signal processor reduces signal fluctuation to ensure the accuracy and stability of data.

      ?  Highly precise peristaltic pump extends the service life of pump tubes.

      ?  Automatic restart when normal power supply after accidental power down.

      ?  Remote control with local or remote PC connection.

      ?  Low reagent consumption; short preparation time.

      ?  Automatic diagnosis function. Identification and alarm for reagent shortage.

      ?  Instrument operating signal output function.

      ?  3000 groups of data storage and USB download for analysis data and running log.

      ?  Easy operation, no special training is required.

      ?  Controllable running of water pump and free setup for the sampling time of the pump.

      ?  Security management function, automatic record and storage for all control operations. Read-only storage.

      ?  2-level operation & management authorization: The system administrator can have all system operation and general operators can only perform routine maintenance & operation.


      Measuring Principle

      NH3-N: The sample after proper filtration is pumped into the reactor, where R1(phenol, sodium nitroprusside, potassium tartrate and potassium hydroxide) and R2 (sodium dichloro isocyanurate) are injected and heated to 50°C. Sample reacts with phenol, potassium tartrate and sodium dichloro isocyanurate under alkaline conditions (using sodium nitroprusside as a catalyst ). After fully mixing and reaction, the analyzer measures ABS value of blue substances produced by reaction at the specific wavelength, then calculates sample concentration according to the stored calibration data.

          PO4: The sample after proper filtration is pumped into the reactor, where molybdate and antimony potassium tartrate are injected. Then, injecting ascorbic acid into the sample. After fully mixing and reaction, the analyzer measures ABS value of blue substances produced by reaction at the specific wavelength, then calculates sample concentration according to the stored calibration data.

          TP: The sample after proper filtration is pumped into the reactor, where injected with acid reagent and heated to 95°C, to convert all inorganic phosphides into orthophosphate. Through adding oxidation reagents and passing through a UV digestion step, to convert the rest of organic forms of phosphide into orthophosphate. Then, injecting antimony potassium tartrate, sodium molybdate and ascorbic acid into the sample. After fully mixing and reaction, the analyzer measures ABS value of blue substances produced by reaction at the specific wavelength, then calculates sample concentration according to the stored calibration data.


      Specifications

      Measuring Principle

      NH3-N

      PO4

      TP

      Colorimetric, Phenate Method

      Colorimetric, Molybdate Ascorbic Acid   Method

      High Temperature Oxidation & UV   Digestion, Colorimetric,  Molybdate   Ascorbic Acid Method

      Operating Management

      Security management and 2-level   operation management function

      Measuring Type

      Manual or automatic with interval

      Measuring Interval

      Programmable; reverse control   available by RS232

      Measuring Time

      12 minutes

      6 minutes

      30 minutes

      Measuring Range

      0.01~0.5/5mg/L, other ranges available   on request

      Calibration Type

      Manual or automatic with interval

      Precision

      ≤ 5%

      Zero Drift

      5%F.S

      Accuracy

      ±5%

      Range Drift

      5%F.S

      Signal Output

      Standard 4-20mA analog output,

      Standard RS232/RS485 digital output,   MODBUS Protocol

      Alarm

      High limit alarm, low limit alarm,   reagent shortage alarm, calibration error alarm (standard)

      Sample and Waste Delivery

      Pressure free

      Reagent Temperature

      10-30°C

      Reagent Replacement

      3~4 weeks (depending on the ambient   temperature)

      Ambient temperature

      5-40°C

      Protection

      IP55

      MTBF

      720h

      Display

      Color touch screen

      Power supply

      220V AC±10%; power 90W

      Weight

      60kg (without reagents)

      Dimension

      1600×500×400mm(H×W×D) (with the height   of base)


      Optional Functions

      Dual-channel

      One analyzer to monitor on-line for 2 outlets.

      Reagent Cooling

      Longer reagent service life and less reagent replacing.

      Alarm Function

      DI water low limit alarm and waste liquid high limit alarm function.

      Automatic Dilution Function

      Automatic dilution function for high concentration sample.

      Mini Printer

      Print data after measurement.


        Tags:On-line Analyzer for Ammonia and Phosphate and Total Phosphorus Monitoring